Repeatability of Corneal Epithelial Thickness Map Using Anterior-Segment Optical Coherence Tomography in Normal and Corneal Disease Patients
Article information
Abstract
Purpose
To evaluate the repeatability of corneal epithelial thickness (CET) measurements in normal eyes and eyes diagnosed with corneal disease using the epithelial thickness map (ETM) of anterior-segment optical coherence tomography (OCT).
Methods
In this retrospective study, patients with three OCT scans using the ETM mode of Cirrus OCT between October 2021 and January 2024 were reviewed. Two groups of subjects were included: (1) normal subjects with no history of ophthalmic surgery, corneal diseases, and topical antiglaucoma medication uses; and (2) subjects with corneal diseases including dry eye syndrome, recurrent corneal erosion, pterygium, and others. A total of 57 eyes of 57 normal subjects and 106 eyes of 76 patients with corneal disease were included. ETM was analyzed in 25 zones (one zone within 0–2 mm diameter, eight zones within 2–5 mm diameter, eight zones within 5–7 mm diameter, and eight zones within 7–9 mm diameter). Repeatability was evaluated by calculating intraclass correlation coefficient (ICC), coefficient of variation (CoV), within-subject standard deviation (Sw), and Bland-Altman plot.
Results
Among a total of 25 sectors, the normal eyes showed high repeatability (ICC, >0.75; CoV, 2.160%–5.292%; Sw, 0.760–1.653 μm) in 23 sectors, and corneal diseases patients also showed high repeatability (ICC, >0.75; CoV, 4.167%–9.606%; Sw, 1.298–3.340 μm) in 22 sectors. However, the wide range of 95% limit of agreement width of Bland-Altman plot presented in corneal disease group and some peripheral zones in normal eyes indicates some variability of CET measurements.
Conclusions
Except for a few peripheral sectors, ETM of Cirrus OCT provides repeatable CET measurements in normal eyes; however in corneal disease group, repeatability was not consistently high. To measure CET accurately, performing multiple measurements is advised especially in patients with corneal disease and patients in whom peripheral CET values.
The corneal epithelium plays a crucial role in maintaining optical clarity and visual acuity, providing essential defense and refractive functions [1–3]. Its thickness can change by remodeling process of the corneal epithelium in response to compensation of stromal alterations and is influenced by various factors and pathologic conditions such as laser refractive surgery, keratoconus (KC), dry eye syndrome, contact lens wear, age, sex, and topical antiglaucoma medication [1–5].
In order to measure corneal epithelial thickness (CET) accurately, high-frequency scanning ultrasound biomicroscopy (HF-UBM), confocal microscopy, and anterior-segment optical coherence tomography (AS-OCT) are used [6]. HF-UBM provides high resolution images unobstructed by optically opaque intervening ocular structures; however, it requires direct contact with the eye, which can be uncomfortable and necessitate greater effort by the examiner and a higher degree of compliance from the patient [7]. Confocal microscopy provides very high axial resolution, but it is unable to provide a holistic view of the cornea to produce topographic mapping with current technology [8].
Topographic mapping of the CET by using OCT was recently developed. OCT makes it possible to measure CET quickly and conveniently without direct contact [4,9–12]. According to previous studies, CET measurements made by using RTVue (OptovueA) topographic mapping and iVue (Optovue) provide high repeatability [13,14]. However, the repeatability of epithelial thickness map (ETM) scans of another the widely used Cirrus OCT (Carl Zeiss Meditec) was not been evaluated. Therefore, the aim of this study is to evaluate the repeatability of CET measurements obtained using the ETM scan of Cirrus OCT in both normal eyes and eyes with corneal disease.
Materials and Methods
Ethics statement
This study was approved by the Institutional Review Board of Ewha Womans University Mokdong Hospital (No. 2024-09-017-001). The requirement for informed consent was waived due to the retrospective nature of the study. The study adhered to the tenets of the Declaration of Helsinki throughout the study.
Subjects
This is a single-center, retrospective, observational study. Patients who visited the hospital and whose CET values were collected using the ETM mode of the Cirrus OCT between October 2021 and January 2024 were reviewed. Two groups of subjects were included: (1) patients with normal eyes; and (2) patients with corneal abnormalities. All patients were 18 years of age or older and were able to complete the required examinations.
Exclusion criteria in the normal group were a history of ocular surgery (laser refractive surgery, cataract surgery, vitrectomy, corneal transplant surgery, etc.), usage of topical antiglaucoma medications, history or current diagnosis of corneal disease, and inability to complete three AS-OCT scans. One eye per individual was included in the normal group. If both eyes qualified, the right eye was selected.
Patients with corneal abnormalities were included if they had a history or current diagnosis of dry eye syndrome, recurrent corneal erosion, pterygium, or another condition. Also, patients with a history of ocular surgery, patients with a history of topical antiglaucoma medication use, and those unable to complete the three AS-OCT scans were excluded. In the patient group, if both eyes qualified, then both were included in the study.
Imaging device and measurement technique
CET measurements were conducted using the Cirrus OCT, which utilizes a noninvasive ETM mode. Each eye underwent three consecutive measurements, with a total of 25 corneal zones assessed for repeatability (Fig. 1A, 1B).
For epithelial thickness mapping, the anterior segment module attachment was employed. Patients were positioned with their gaze fixed on a target light, and the imaging was centered on the pupil center. The pachymetry map consisted of eight radial scans (each with 1,024 axial scans) repeated five times, covering an area with a 9 mm diameter. The software algorithm calculated CET as the distance between the middle of the tear film layer and the middle of the anterior surface of the Bowman layer observed on the B scan. Images were captured after aligning the horizontal single scan line with the corneal apex, ensuring the visibility of the hyperreflective corneal reflex. Additional scans were performed if the initial scan was misaligned or showed poor reflection of the corneal apex. Data were exported and analyzed using Cirrus OCT Review Software ver. 11.0 (Carl Zeiss Meditec), which provided average automated measurements of CET for four concentric ring-shaped zones centered on the cornea: central (0–2 mm), paracentral (2–5 mm), mid-peripheral (5–7 mm), and peripheral (7–9 mm). ET data were also presented for specific corneal octants: superior (S), inferior (I), temporal (T), nasal (N), superior nasal (SN), superior temporal (ST), inferior temporal (IT), and inferior nasal (IN) within the paracentral, mid-peripheral, and peripheral zones [15].
Statistical analysis
Repeatability was evaluated using the intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), coefficient of variation (CoV), and coefficient of repeatability (CR). The ICC serves as an indicator of the reliability of repeated measurements, and it was calculated using the two-way mixed model and absolute agreement. Its values less than 0.5, 0.5 to less than 0.75, and 0.75 or greater represent poor, moderate, and good repeatability, respectively. To compare ICC value between corneal disease group and normal group, resampling procedure called bootstrap method was used. 1,000 bootstrap samples were generated and 95% coefficient interval (CI) of ICC difference that does not include 0 was considered statistically significant. Sw allows the estimation of variance. The CoV was calculated as the Sw divided by the mean of the measurements (CoV = Sw / mean × 100%). The CoV allows a comparison between data that have different means. A larger CoV reflects a greater level of dispersion around the mean. Bland-Altman plots were used to analysis the distribution of variability by size, mean difference and 95% limit of agreement (LoA), and CR of three pairs, which were formed by combining two out of the three measurements: pair 1 (measurement 1 – measurement 2), pair 2 (measurement 2 – measurement 3), and pair 3 (measurement 3 – measurement 1). All statistical analyses were performed using IBM SPSS ver. 26.0 (IBM Corp), R ver. 4.4.1 (R Core Team), and MedCalc ver. 23.0.5 (MedCalc Software).
Results
The normal group consisted of 57 eyes of 57 patients (30 men and 27 women) with a mean age of 52.6 ± 17.1 years, while the corneal disease group included 106 eyes of 76 patients (30 men and 46 women), with a mean age of 50.1 ± 15.5 years. Enrolled subjects with corneal pathology had one of the diagnoses listed in Table 1.
The average CET and repeatability factor by diameter in normal and corneal disease eyes are presented in Table 2. In the normal eyes group, the average CET for the 0–2 mm diameter zone was 48.34 ± 3.73 μm, with an ICC of 0.918, Sw of 0.827 μm, CoV of 2.214%, and CR of 2.941 μm. For the 2–5 mm diameter zone, the average CET was 46.62 ± 4.23 μm, ICC was 0.875, Sw was 1.013 μm, CoV was 2.973%, and CR was 3.803 μm. In the 5–7 mm zone, the average CET was 44.47 ± 4.68 μm, ICC was 0.866, Sw was 1.136 μm, CoV was 3.598%, and CR was 4.229 μm. Finally, in the peripheral 7–9 mm zone, the average CET was 43.46 ± 5.60 μm, ICC was 0.793, Sw was 1.571 μm, CoV was 4.720%, and CR was 5.775 μm. For the corneal disease group, the average CET in the 0–2 mm diameter zone was 50.15 ± 7.48 μm, with a n ICC of 0.939, Sw of 1.298 μm, CoV of 5.993%, and CR of 5.076 μm. In the 2–5 mm zone, the average CET was 48.43 ± 6.89 μm, ICC was 0.902, Sw was 1.587 μm, CoV was 5.816%, and CR was 5.972 μm. For the 5–7 mm zone, the average CET was 46.19 ± 6.24 μm, ICC was 0.845, Sw was 1.767 μm, CoV was 7.047%, and CR was 6.918 μm. In the 7–9 mm zone, the average CET was 45.05 ± 6.53 μm, ICC was 0.744, Sw was 2.383 μm, CoV was 9.622%, and CR was 9.097 μm.
Fig. 2A–2D and Table 3 shows CET and repeatability index by sectors of normal and corneal disease patients. Out of 25 sectors, the normal eyes showed high repeatability (ICC, >0.75; CoV, 2.160%–5.292%; Sw, 0.760–1.653 μm) in 23 sectors except for 2 sectors (7–9 mm S and IT), and patients with corneal diseases also showed high repeatability (ICC, >0.75; CoV, 4.167%–9.606%; Sw, 1.298–3.340 μm) in 22 sectors, except for 3 sectors (7–9 mm SN, IN, and IT). The 95% CI of the ICC difference between two groups, calculated after 1,000 of resampling using the bootstrap method, is also presented in Table 3. There was no statistically significant difference between ICC values of normal and corneal disease groups, except in three sectors, 5–7 mm T, 7–9 mm N, and 7–9 mm T, although these three zones are neither significant regions nor regions with low repeatability indicators.
Fig. 3A–3F and Supplementary Fig. 1–3 show Bland-Altman plots of three pairs at every sector in normal and corneal disease patients. Tables 4 and 5 present the repeatability metrics analyzed using the Bland-Altman plot in normal and corneal disease patients. For the normal group, the mean difference ranged from –0.96 to 0.67 μm, 95% LoA width ranged from 4.552 to 22.074 μm, and the CR ranged from 2.276 to 11.461 μm. For the corneal disease group, the mean difference ranged from –0.90 to 0.78 μm, 95% LoA width ranged from 7.080 to 27.438 μm, and the CR ranged from 3.540 to 13.719 μm. In both normal and corneal disease groups, the absolute value of the mean difference was less than 1 μm, however, wide 95% LoA range were found in corneal disease eyes and peripheral zones in both groups.
Overall, ICC values in both groups indicated high repeatability in the central to mid-peripheral zones (0–2, 2–5, and 5–7 mm), while peripheral zones (7–9 mm) showed reduced reliability, particularly in the corneal disease group. Sw, CoV, and CR values show increasing trends toward the periphery, indicating greater variability in these regions. This variability was more pronounced in the corneal disease group, where CoV exceeded 9% in the 7–9 mm zone, suggesting less consistent measurements and greater individual variation compared to the normal eyes group. In both groups, repeatability and measurement rate showed a decreasing trend as the diameter increased. The measurement rate was nearly 100% for diameters up to 7 mm, but the average measurement rate for the 7–9 mm diameter was 93% (ranging from 81% to 100%) in the normal eyes group and 88% (ranging from 66% to 100%) in the corneal disease group.
Discussion
The aim of this study was to determine the repeatability of CET measurements with the ETM of AS-OCT (Cirrus OCT) in normal and diseased corneas. We evaluated repeatability using various methods, including ICC, Sw, CoV, and Bland-Altman plot. Even though good repeatability was observed in both groups except for some peripheral zones between 7 and 9 mm in ICC, CoV, and Sw values, wide range of 95% LoA of Bland-Altman plot indicated considerable variability in the corneal disease group. Furthermore, both the normal and corneal disease groups exhibited a decreasing trend in repeatability metrics and measurement rates as the diameter extended toward the periphery.
Several previous studies have evaluated the repeatability of ETM using AS-OCT in normal eyes by measuring key statistical parameters (ICC, Sw, and CoV), which provide objective measures of the consistency and reliability of ETM measurements across different devices and ocular conditions. Kanellopoulos and Asimellis [16] used RTVue SD-OCT, which employs a scan of 17 sectors (one sector for 0–2 mm, eight sectors for 2–5 mm, and eight sectors for 5–6 mm), with 373 normal subjects and reported Sw 0.88 ± 0.71 μm in the center. Hashmani et al. [17] used wide 25 sectors, 9-mm Optovue SD-OCT with 220 normal subjects, and they reported ICCs of 0.703 to 0.812 at the inner circle, 0.712 to 0.917 at the middle circle, and 0.796 to 0.930 at the outer circle, which indicates good repeatability in most zones except a few sectors (center, inner IN, I, IT, and middle IN), and there was no significant difference between the repeatability of the center, paracentral, and periphery. Sikorski [18] used REVO NX (Optopol Technology), 17 sectors with 7-mm diameter, on 137 patients with normal eyes, and the ICC was 0.95 at the center (0–2 mm), 0.82 to 0.91 at paracentral (2–5 mm), and 0.64 to 0.89 at peripheral (5–7 mm); they noted high repeatability in most areas except for some at the periphery (5–7 mm S, T, IT, SN), and the repeatability value diminished toward the periphery. Sella et al. [4] reported good corneal ETM repeatability (standard deviation: 0.9 μm at 0–2 mm, 0.9–1.3 μm at 2–5 mm, 1.0–1.4 μm at 5–6 mm), and values decreased toward peripheral in 12 normal eyes across all zones with iVue, which provides scans of 17 sectors (one sector for 0–2 mm, eight sectors for 2–5 mm, and eight sectors for 5–6 mm). Similar to the above studies, except for one study by Hashmani et al. [17] that reported no significant difference between the center and the periphery, our study also showed that repeatability values tend to decrease towards the periphery in all repeatability indices including ICC, Sw, CoV, and CR values. Reinstein et al. [19] suggested that the reason for the decline in the signal quality of CET measurements taken further from the center is that as measurements shift toward the periphery, the OCT beam fails to strike the cornea at optimal angles, resulting in diminished reflection and inadequate signal production.
It is essential to understand normal values and variation patterns in CET for interpreting the information provided by ETM, which has extensive clinical applications by enhancing diagnosis and management of corneal disorders like KC, corneal dystrophies, limbal stem cell deficiency, and dry eye disease [20,21]. Furthermore, the ETM improves refractive surgery safety and efficacy through better patient screening, procedure planning, and postoperative monitoring [21,22]. Analysis of ETM data and pattern alterations may help diagnosing and staging of KC and evaluating the efficacy of collagen crosslinking and combined procedures [23,24].
Previous studies have evaluated the repeatability of CET measurements using the ETM of AS-OCT in patients with corneal diseases, and most were conducted using RT-Vue— except for one by Li et al. [25]. Ma et al. [13] reported good repeatability with an ICC of 0.891 or higher in all regions for 45 post–laser-assisted in situ keratomileusis (post-LASIK) eyes. Mohr et al. [26] reported good repeatability with an ICC ranging from 0.841 to 0.981 in all regions for 59 KC eyes and also observed a trend toward reduced repeatability value with increased distance from the center. Lu et al. [27] investigated 68 post–photorefractive keratectomy eyes (CoV, 2.6%–6.2%), 61 post–small incision lenticule extraction (post-SMILE) eyes (CoV, 2.3%–4.7%), 75 post–femtosecond LASIK eyes (CoV, 4.0%–6.3%), 20 mild KC eyes (CoV, 2.5%–6.2%), and 53 advanced KC eyes (CoV, 3.5%–8.0%); CoV values were higher in advanced KC compared to mild KC patients. Li et al. [25] used REVO NX, which employs a scan of nine sectors (center, S, ST, T, IT, I, IN, N, SN) at 8-mm diameter, with 259 KC eyes and reported good repeatability in all sectors (ICC, 0.82–0.95; CoV, 3.20%–5.46%). Consistent with the previous studies, the present study also showed good repeatability indices of CET measurements using ETM in patients with corneal diseases in 22 sectors, ICC of 0.768 to 0.939, CoV of 4.17% to 9.61%, and Sw of 1.29 to 3.34 μm, except for 7–9 mm sectors of SN (ICC, 0.669; CoV, 14.12%; Sw, 2.762 μm), IN (ICC, 0.644; CoV, 11.76%; Sw, 3.241 μm), and IT (ICC, 0.586; CoV, 13.36%; Sw, 3.340 μm). Moreover, there was no statistically significant difference between two groups when comparing the ICC values using the bootstrap method. However, the wide 95% LoA width of Bland-Altman plot in all zones in corneal disease showed considerable reliability in these patients. Further investigation involving large cohort of patients with various corneal disease is needed to achieve consistent repeatability indices and conclude the repeatability of CET measurements with ETM of Cirrus OCT in corneal disease patients.
The average center CET of the normal group in this study was 48.3 ± 3.7 μm, similar to a study by Loureiro et al. [20] that reported center CET in 20 normal eye measurements by using Cirrus OCT to be 47.9 ± 1.2 μm. Unlike the results using Cirrus OCT, center CET measurements of normal subjects using RTVue reported in other studies are 52.8 ± 3.6 μm [5], 53.3 ± 3.3 μm [16], 52.3 ± 3.6 μm [28], and 50.5 ± 3.9 μm [29]. When comparing the center CET measurements of Cirrus OCT with RTVue center CET value, OCT with Cirrus was approximately 4 to 5 μm smaller than in RTVue devices, which is in agreement with Loureiro et al. [20]. The Cirrus OCT measures CET as the distance from the midpoint of the tear film to the midpoint of the anterior surface of the Bowman layer, which excludes some portion of the tear film [15]. In contrast, RT-Vue and iVue include the tear film in their measurements. Tear film is known to be about 2 to 5.5 μm [21,23,30], and since this value is similar to the difference in thickness seen in studies, this CET measurement difference can be estimated by including the tear film layer in OCT measurement.
In conclusion, Cirrus OCT ETM can be used reliably for CET measurement in normal eyes, but in corneal disease eyes, repeatability indices were not consistently high, making it advisable to perform multiple measurements to confirm the CET values. The observed variability in peripheral zones is consistent with previous studies using other OCT devices, suggesting inherent challenges in measuring these areas. This study offers foundational data for accurate CET assessment. The high repeatability in central zones underscores the potential of the Cirrus OCT as a valuable tool in clinical settings for diagnosing and monitoring corneal health. Further studies could explore strategies to enhance measurement reliability in peripheral regions.
Acknowledgements
None.
Notes
Conflicts of Interest:
None.
Funding:
None.
Supplementary Materials
Supplementary Fig. 1. Bland-Altman plots of three pairs in each sector of 2–5 mm diameter in normal and corneal disease eyes.
kjo-2024-0115-Supplementary-Fig-1.pdfSupplementary Fig. 2. Bland-Altman plots of three pairs in each sector of 5–7 mm diameter in normal and corneal disease eyes.
kjo-2024-0115-Supplementary-Fig-2.pdfSupplementary Fig. 3. Bland-Altman plots of three pairs in each sector of 7–9 mm diameter in normal and corneal disease eyes.
kjo-2024-0115-Supplementary-Fig-3.pdfSupplementary materials are available from https://doi.org/10.3341/kjo.2024.0115.